Page 29 - FoodFocusThailand No.161 August 2019
P. 29

SPECIAL FOCUS


                     Accuracy Accuracy is the closeness of test results   Limit of Detection (LOD) and Limit of Quantitation (LOQ)
                  1obtained by test method to the true value. Certified  4LOD and LOQ are two important performance characteristics in method
                  Reference Material (CRM) is highest recommended to be   validation, especially for analysis of trace amount of food toxicants/contaminants.
                  analyzed at least 7 determinations whenever possible   LOD is the smallest concentration of analyte that can be detected, but not necessarily
                  because their highest level of traceability. In case of CRM   quantified. While LOQ is the smallest quantified concentration of analyte in the
                  is limited, spiked sample method can be a good representation,   sample that can be determined with acceptable precision and accuracy. This
                  at least 3 spiked concentrations (with 7 determinations for   approach could only be applied to analytical procedure which exhibits baseline
                  each concentration) are recommended.          noise by determination of Signal-to-Noise ratio (S/N). LOD and LOQ are generally
                                                                expressed as 3 and 10 times of standard deviation of response of blank sample or
                     Precision Precision is the determination of closeness   with known low concentrations, respectively.
                  2of agreement replicated and independent test results,
                                                                  Ruggedness  Ruggedness  is  a parameter  expressed  the degree  of
                  obtained under stipulated conditions. Repeatability is the   5reproducibility of test results obtained by the analysis of the same samples
                  precision obtained under “same test condition”, same
                  laboratory, same operator, same equipment. While   under a variety of conditions, such as different elapsed testing time, different analyst,
                                                                different instrument, different pH of water, or different assay temperature.
                  reproducibility is another precision parameter obtained when
                  test results are obtained with the same method, but “under  Additional Guidelines for Method Validation
                  changing test condition” e.g. different laboratory, different   When the laboratory have used a fully validated method, but new food matrix(s)
                  operator,  or  different  equipment.  Precision  is  usually   are applied, or in situation that the single laboratory has developed fully-validated
                  expressed in terms of Standard Deviation (SD) or Coefficient   method, but no collaborative study has not yet applied; in these cases, the laboratory
                  of Variation (CV) of the test results.        needs to extend the verification of accuracy, precision, linearity and matrix effect to
                                                                ensure the appropriate of test results. When the laboratory needs to use the test
                     Linearity and Range Linearity is its specification   method that published in the scientific literature, however no characteristics given,
                  3 of test method to obtain test results which are directly   in this case, laboratory needs to verify accuracy, precision, linearity as well as
                  proportional to the concentration/amount of analyte in the   ruggedness. It is essential that appropriate validation (at least accuracy and
                  sample. A linear relationship shall be evaluated across the   precision) is always be carried out before taking any new method into routine use.
                  range by visual inspection of a plot of signals as a function   Verification is recommended to repeat in cases such as, when major instruments
                  of analyte concentration. For the establishment of linearity   are replaced, new batches of major reagents are used, or even changing the
                  of any test, a minimum of 5 concentrations is recommended.   operators.
                  While range of an analytical procedure is the interval between
                  the upper and lower of analytes concentration in the food   เอกสารอ้างอิง/Reference
                  matrix for which it has been demonstrated that the analytical  Eurachem Guide. 2014. The Fitness for Purpose of Analytical Methods, A Laboratory Guide to
                  procedure has a suitable level of precision, accuracy and    Method Validation and Related Topic, 2nd edition.
                  linearity.














 Method Validation Guidelines

 for Chemical Analysis of Food Matrix

 as Requirement of ISO/IEC17025
























                                                                                                  AUG  2019 FOOD FOCUS THAILAND  29


         27-29_Special Focus_Kanitaporn.indd   29                                                                    20/7/2562 BE   14:02
   24   25   26   27   28   29   30   31   32   33   34